10-32
Graphics Menu
Single Graph
displays the current vs. potential plot(s) of the
Data Sets
specified in the
Graph
and
Results
Options
dialog boxes.
Analysis Menu
The
Auto
option for
Results Graph
displays the current vs. potential plot(s)
of the
Data Sets
specified in the
Graph
and
Results Options
dialog boxes,
and the peak potential and peak current are listed in the
Main
window.
Alternative baselines can be set by the user through the
Manual
option.
Mode = BSWV/P
General Parameters
Initial E (mV) = 0 -3276 to 3276
Final E (mV) = 0 -3276 to 3276
Sensitivity = 1
µ
A/V 100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Specific Parameters
Scan Rate (mV/s) = 4 1 to 200
S.W. Amplitude (mV) = 25 1 to 250
S.W. Frequency (Hz) = 14 5 to 250
Drop Time (msec) = 1500 1500 to 8000
Quiet time (sec) = 2 (0) 0 to 65535
Comments
a) Parameters values in parentheses refer to
BSWP
. The
Drop Time
is only
used for
BSWP
(the analogous parameter for
BSWV
, the
Step Period
, is set
at 1500 msec).
b) The number of current samples averaged during each
Drop Time
(or
Step
Period
) is determined by the
S.W. Frequency
.
c) Sampling width is fixed at 1/4 cycle time period for all frequencies.
d) The sample interval is the product of the
Scan Rate
and the
Drop Time
(or
Step Period
) (e.g., the default value is 6 mV).
Graphics Menu
Single Graph
displays the difference current vs. potential plot. The forward
and reverse current plots are not available for
BSWV/P
.