4-4
4.6
Deposit Options
Deposit Options
is used to define the deposition potential for stripping experiments.
The default for this parameter is the initial potential for the potential scan (
Initial E
).
If some other potential is required, then the
Deposit E
option should be selected, and
the required potential should be entered in the
Deposit E
field (Figure 4-5).
Figure 4-5. Deposit Options
dialog box.
4.7
Stabilizing Capacitor
A stabilizing capacitor is only required when using iR compensation, or for Bulk
Electrolysis (
BE
) experiments. In both these cases, the
Auto
option switches a 0.1
µ
F
or 1
µ
F capacitor into the circuit. The
Manual
option generally only needs to be used
when using the
Manual iR
option (Figure 4-6). The
Manual
option should be used
with caution, since the addition of capacitors can affect the frequency response of the
system.
Figure 4-6. Stabilizing Capacitor
dialog box.