7-1
7. Analysis
7.1
Results Graph
The data available from an electrochemical experiment may be peak potentials and
currents, half-wave potentials and limiting currents, slopes and intercepts of straight
line plots or the amount of charge passed during the experiments. These data can be
obtained using the
Results Graph
.
There are two options available:
Auto
and
Manual
. These can be selected from the
Results Options
dialog box. When the
Auto
option is selected, the
Results Graph
windows displays the experimental data, the baseline for peak measurements and the
best fit straight line for linear plots (the size and position of this windows can be
changed by the user (by dragging the borders or the caption), and are defined initially
by the BAS100W.INI file). The
Auto
option also selects the appropriate
Peak Shape
.
The experimental results (see below) are listed in the
Main
window.
CV
,
LSV
Selected plot - tailed curve
Experimental results - peak potential, current and
area
HR (even-numbered
Selected plot - tailed curve
segments)
Experimental results - peak potential, current and
area
ACV/P
,
BSWV/P
,
DPV/P
, Selected plot - symmetric curve
DTACV/P
,
OSWV
, Experimental results - peak potential, current and
area
PSACV/P
,
TACV/P
BSWSV
,
DPSV
,
LSSV
, Selected plot - symmetric curve
OSWSV
Experimental results - peak potential, current and
area
HDM
,
NPV/P
,
RDE
,
SCP
Selected plot - sigmoidal curve
Experimental results - half-wave potential
CA, CC
Selected plot - straight line plots (i vs. t
-
1/2
for CA,
Q vs. t
1/2
for CC)