10-66
10.9.3
Hold-Ramp-Step (HR)
HR
allows the programming of up to 12 time and/or potential based segments. It is
used to build custom waveforms for special applications.
Odd-numbered segments are chronoamperometric (current vs. time at a fixed
potential) and even-numbered segments are voltammetric (current vs. applied
potential ramp).
Mode = HR
General Parameters
Time Domain = (sec,
µ
V/sec) or (msec, mV/sec)
Data Set = 1, 3, 5, 7, 9, 11
Hold E (mV) = 0 -3276 to 3276
Hold Time = 0 0 to 65535
Data Sets 2, 4, 6, 8, 10, 12
Initial E (mV) = 0 -3276 to 3276
Final E (mV) = 0 -3276 to 3276
Scan Rate = 100 1 to 51200
Sensitivity = 1
µ
A/V 100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Specific Parameters
Quiet Time (sec) = 2 0 to 65535
Graphics Menu
Single Graph
displays the current vs. time plot for the
Data Set
selected
using
Data Sets
in
Graphics Options
.
Analysis Menu
The
Auto
option for
Results Graph
displays the current vs. potential curve
for the even-numbered Data Set selected using
Data Sets
in
Graphics
or
Results Options
.