10-71
Mode = IMP
General Parameters
Initial E (mV) = 0 -3276 to 3276
Low Frequency (Hz) = 1E 1 -4 to 2
High Frequency (Hz) = 1E 3 -2 to 3
A.C. Amplitude (mV) = 5 1 to 250
Sensitivity = Auto Auto, 100 mA/V to 1
µ
A/V
Specific Parameters
Freq. Increment (decade) = 1 1 or 2
100-1000 Hz = 64 1 to 400
10-100 Hz = 16 1 to 400
1-10 Hz = 16 1 to 400
0.1-1 Hz = 8 1 to 400
0.01-1 Hz = 8 1 to 400
0.01-0.1 = 2 1 to 400
0.001-0.1 = 2 1 to 400
0.001-0.01 = 1 1 to 400
0.0001-0.01 = 1 1 to 400
Quiet Time (sec) = 2 0 to 65535
Comments
a) The
A.C. Amplitude
should be 8/n mV or less (n = number of electrons
transferred).
b) For the two lowest frequency decades, each cycle requires ca. 3 hrs.
Therefore, no more than one cycle should be used since the system may
change significantly over the time period of the experiment. If the system is
relatively unstable, it may not be possible to use these decades at all.