Profiling
You can define the treatment parameter pattern through profiling. Ultrafiltration rate (UF rate), sodium (Na
+
)
concentration, and bicarbonate (HCO
3
-
) concentration profiles can be used by the machine. You must confirm the
profiling parameters prior to treatment.
Profiling Settings/Activation
You may set up the profiling parameters immediately after the machine turns on and Function Check appears on
the machine’s status indicator. You may set up a model manually or select a preset model. You may set/adjust
profiles for sodium (Na
+
) concentration, bicarbonate (HCO
3
-
) concentration, and UF rate for each model. The UF
goal must be set up for treatment in order to use the ultrafiltration profile.
Note: If UF goal or treatment time changes during active UF profiling, the profile parameters will change
automatically. Always check the UF profile settings if UF goal or treatment time changes when a UF
profile is activated. If the profile parameters have not changed and UF profiling reactivates, the UF
profiling graph will change automatically according to the remaining UF goal and treatment time.
Always check the UF profile settings after deactivating and reactivating a UF profile.
Ultrafiltration Profiling
1. Press the Fluid button.
2. Select the UF rate tab.
3. Press Profiling.
4. Press UF rate.
5. Press Mode.
6. Select the graph method (linear, step, or interval). If the profile is
step or interval, select the number of steps or intervals according to
the prescription.
7. Press OK.
8. Select the Start value for the UF rate. The Stop value for the UF rate
will be automatically calculated in relation to the starting value.
9. Press Activate.
10. Change or set the profiles for Na
+
or HCO
3
-
, if applicable.
Note: You can create and use models by having an authorized
technician preset the values for each profile.
Profiling
ADVANCED FUNCTIONS OF AK 98
29

Table of Contents

Other manuals for baxter AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter AK 98 and is the answer not in the manual?

baxter AK 98 Specifications

General IconGeneral
Brandbaxter
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals