Manually Entering the Prescription
Set Treatment Time
1. Press the time indicator on the screen.
2. Press treatment time.
3. Set the desired treatment time and press OK.
Set Ultrafiltration Goal
4. Press the time indicator on the screen.
5. Press UF volume.
6. Enter the patient’s ultrafiltration goal and press OK.
Check the calculated/resulting UF rate.
Set Heparin Values
7. Press the Blood button.
8. Select the Heparin tab.
9. Press the settings one at a time, then set up the desired
parameters, such as heparin bolus dose, heparin flow rate,
stop time, etc. according to the prescription.
10. If a heparin bolus dose is needed at the start of treatment,
press Heparin bolus dose and set a bolus dose (mL)
appropriate for the patient.