To avoid unnecessary air detector alarms during set up, wait to activate the air detector until all air has been removed
from the circuit.
Note:
The air detector is automatically activated when Connect patient is selected.
The “Flashing” Attention Hand
Alarms have higher priority than attention messages on the
screen. This means that if an alarm occurs and cannot be resolved,
discontinuing the treatment may only be possible by pressing the
flashing attention hand on the screen.
Once the flashing attention hand is selected, the “Confirm treatment
time expired” message will be accessible.
Note:
It may be necessary to repeat this sequence to complete the
treatment discontinuation procedure.
Air Detector Activation
The actual value is the current value seen on the operator’s panel.
The calculated value is based on the type of concentrate selected and
the set values for sodium and bicarbonate.
Note:
If the two values do not match, the dialysate will
automatically bypass the dialyzer.
Conductivity Actual and Calculated - C/P Values
TIPS FOR USING AK 98
Tips to support accurate heparin delivery:
Ensure the syringe used is the same as the brand and size
programmed into the AK 98 presets, or the heparin delivery may not
be accurate.
Confirm that heparin line is fully primed.
To view the syringe brand and size, go to Blood Menu and select
Heparin.
Heparin Delivery
If restarting using the On/Off button located on the operator’s panel,
to the right of the screen, ensure the dialysate lines are properly
connected to the color-coded standby ports on the machine before
restarting the AK 98. This will avoid unnecessary error messages.
Restart Using the On/Off Button