35
34
AK 98 can be placed into concentrate standby mode. In concentrate standby mode, the concentrate is no longer
used, and water consumption can be reduced. Concentrate standby mode can be activated when the functional
check is done and the flow diagram turns green. When concentrate standby mode is active, a message will appear
in the information field. The bypass path of the flow diagram will turn orange. After 1 hour in concentrate standby
mode, the machine will automatically start preparing dialysate again. If blood is detected, concentrate standby
mode cannot be activated. Concentrate standby mode can be activated manually, or preset to activate automatically.
Recirculation and concentrate standby mode cannot occur simultaneously.
How to Activate Concentrate Standby Mode Manually
1. Press the Fluid button.
Note: The machine can be preset to automatically
activate concentrate standby mode.
2. Select the Dialysate tab.
3. Activate concentrate standby mode.
How to exit Concentrate Standby Mode
1. Select the Information tab.
2. Press Confirm the message.
Note: Do not stop the blood pump to connect patient until
the dialysate line is green.
Concentrate Standby Mode (CSBM)
OTHER FUNCTIONS OF AK 98
CSBM will automatically deactivate after 1 hour.
To reactivate CSBM press the Fluid key followed by the Dialysate
tab and activate CSBM
.
Repeat every hour as needed.
Reactivate Concentrate Standby Mode (CSBM)