Page #16 background imageLoading...
Page #16 background image
17
16
The Patient button is filled in and has a check mark. This confirms the
machine is running with the patient prescription coming from the
EHR/EMR.
Note: The treatment parameters can also be entered manually in the
Patient menu under the Settings tab.
Retrieving Patient Prescription
HOW TO USE AK 98
Note:
Applies only for bidirectional connectivity.
1. Press the Patient button.
2. Press the Patient ID bar.
3. Enter your ID and press OK.
4. Carefully check the correct patient data has been retrieved,
and then press Confirm to accept.
5. Review the prescription dialog box.
Shows the prescription parameters which have come from the
EHR/EMR. The other parameters are either your facility's settings,
or the machine default values.
Any parameter can be changed if needed by pressing the blue bars.
Shows the prescription parameters that have come from the
EHR/EMR and have been changed.
6. Verify that all the required treatment parameters have been
set correctly, and then press OK.

Table of Contents