HOW TO USE AK 98
Disinfection and Rinsing
Press the flashing Disinfection button to disinfect/rinse.
Chemical substances or heat disinfection can be used to disinfect.
Chemical Disinfection
1. Select the “Chemical” tab from the Disinfection/Rinse menu.
2. Select the disinfectant.
3. Connect the blue concentrate connector to the yellow wand.
4. Place the wand inside the disinfectant solution.
5. When the machine has withdrawn the disinfectant solution,
an alarm will sound. Disconnect the concentrate connector
from the wand and connect it to the machine.
Heat Disinfection
1. Select the “Heat” tab from the Disinfection/Rinse menu.
2. Select “Heat”.
3. Press Confirm.
Heat Disinfection With Liquid Citric Acid
1. Make sure the disinfectant inlet tube from behind the machine is
placed in the citric acid solution.
2. Select the “Heat” tab from the Disinfection/Rinse menu.
3. Select “Citric 20%" or "Short heat citric" and press Confirm.
Note: Citric acid concentration is a preset.