HOW TO USE AK 98
Connecting the Patient
Note:
Blood pump must be stopped after recirculation for the Connect
patient option to appear. Wait to stop the blood pump until the
dialysate line on the operator's panel is green.
1. Check that the venous line is correctly placed in the priming detector.
2. Press the Connect Patient button on the screen to enter patient
connection mode.
10. Once the blood passes through the dialyzer and blood is detected
by the priming detector, the blood pump will stop automatically.
11. The color of the blood path will turn red on the screen.
3. Clamp the infusion lines.
4. Clamp arterial and venous blood lines. Disconnect the arterial blood
line from the recirculation connector.
5. Connect the arterial blood line to the arterial access (needle or
catheter). Make sure the connection is tight and that there is no
leakage.
6. Check that there is no air in the venous blood line.
7. Connect the venous blood line to the venous access (needle or
catheter). Make sure the connection is tight and that there is no
leakage.
8. Check that all relevant clamps are closed to prevent blood loss.
9. Press the flashing blood pump button to start the blood pump.
The blood pump will start at the preset flow (default 100 mL/min).

Table of Contents

Other manuals for baxter AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter AK 98 and is the answer not in the manual?

baxter AK 98 Specifications

General IconGeneral
Brandbaxter
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals