ADVANCED USE MANUAL DD700
NOTES
MPP: commands for MPP weights memory management (optional)
Slot1.slot2: optional cards
Note 1): the status variable contains the following indicators:
bit
meaning
15
Center zero signal (-0.25dd < gross < 0.25dd)
14
preset(1)/self-weighing(0) tare signal
13
Blocked tare
12
minimum weighing operation
11
MSB weighing extension signal (only ME)
10
overload
9
Stable weight
8
LSB weighing extension signal (only ME)
7
RFU
6
Invalid weight
5
deleted blocked tare signal (only ME)
4
Entered tare
3
RFU
2
error on scale configuration parameters
1
Failed converter
0
Approved terminal
Note 2): the variable contains some status flags
bit
meaning
15
MPP operation in progress
14
Data ready
13
RFU
12
Invalid weight error
11
Negative weight error
10
Weight not stable error
9
RFU
8
MPP memory error
7
Vehicle positioning error
6-0
RFU
Note 3): the variable is only available with multi scale terminals and contains the number of the scale
selected in the low byte; A=1, B=2,
bit
meaning
15-8
RFU
7-0
Scale number