BD / ED / FD (E2) 11/2010 page 18/54
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
Transport the unit only in its original packaging.
Secure the unit with transport straps for transport.
∅ Do NOT lift or transport the unit using the door handle or the door.
∅ Do NOT lift units size 400 and 720 by hand.
Lift unit size 23, 53, 115 at its four lower corners with the aid of 2 people, unit size 240
with the aid of 4 people, and place it on a transport pallet with wheels. Push the pallet
to the desired site and then lift the unit from the pallet at its four lower corners.
Place units size 400 and 720 using technical devices (fork lifter) on the transport pallet.
Set the fork lifter only from the rear in the middle of the unit. Make sure to place all the
lateral supports of the unit on the forks.
Transport units size 400 and 720 ONLY with the original transport pallet. Set the fork
lifter only to the pallet. Without the pallet the unit is in imminent danger of overturning!!
• Permissible ambient temperature range during transport: -10 °C to +60 °C / 14 °F to 140 °F.
You can order transport packing and pallets for transportation purposes from BINDER Service.
3.3 Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 10.2).
•
Permissible ambient temperature range during storage: -10 °C to +60 °C / 14 °F to 140 °F.
• Permissible ambient humidity: max. 70 % r.H., non-condensing
When after storage in a cold location you transfer the unit to its warmer installation site, condensation
may form. Before start-up, wait at least one hour until the unit has attained ambient temperature and is
completely dry.
3.4 Location of installation and ambient conditions
Set up the incubators BD or the heating oven ED / FD on an even and non-flammable surface, free from
vibration and in a well-ventilated, dry location and align it using a spirit level. The site of installation must
be capable of supporting the unit’s weight (see technical data, chap. 12.4 to 12.6). The chambers are
designed for setting up inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the unit.
∅ Do NOT set up units in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
• Permissible ambient temperature range during operation: +18 °C up to +40 °C / 64.4 °F to 104 °F.
• . At elevated ambient temperature values, fluctuations in temperature can occur.