BD / ED / FD (E2) 11/2010 page 35/54
In case of impurity of the interior with biological or chemical hazardous material, there are three possible
procedures depending on the type of contamination and of the charging material.
1. The heating ovens ED and FD can be hot air sterilized at 190 °C / 374 °F for at least 30 minutes. All
inflammable goods must be removed from the interior before. With the incubators BD it is possible to
perform a hot-air disinfection at 100 °C / 212 °F.
2. Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolute dry and ventilated, because explosive gases may form
during the decontamination process.
3. If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or
have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
With every decontamination method, always use adequate personal safety controls.
CAUTION
Danger of corrosion.
Damage to the unit.
Do NOT use acidic or chlorine cleaning detergents.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
and mild cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
9.3 Sending the unit back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An authorization number
will be issued after receiving your complaint either in writing or by telephone prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
below:
BINDER product type and serial number
Date of purchase
Name and address of the dealer from which you bought the BINDER product
Exact description of the defect or fault
Complete address, contact person and availability of that person
Exact location of the BINDER product in your facility
A contamination clearance certificate (chap. 13) must be faxed in advance
The authorization number must be applied to the packaging
in such a way that it can be easily
recognized or be recorded clearly in the delivery documents.
For safety reasons we cannot accept a unit delivery if it does not carry an authorization
number.