BD / ED / FD (E2) 11/2010 page 33/54
8.5 Water protected internal socket (option BD)
The internal socket is splash proof.
IP system of protection 65 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 50 °C / 122 °F.
WARNING
Exceeding of the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
Do NOT exceed the temperature set-point of 50 °C / 122 °F.
Set the mechanical thermostat class 3.1 to 50 °C / 122 °F.
Heat emission of electrical devices connected inside the chamber may modify the
temperature range.
CAUTION
Risk of short circuit.
Damage to the unit.
Use the delivered plug only (IP protection type 66). Plug-in the plug and turn it to
secure.
If the socket is not used, close the lift-up lid and turn it for securing.
9. Maintenance, cleaning, and service
9.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
The unit must NOT become wet during operation or maintenance work.
Do NOT remove the rear panel of the unit.
Disconnect the unit before conducting maintenance work. Disconnect the power plug.
Ensure all maintenance work is conducted by licensed electricians or experts
authorized by BINDER.
Ensure regular maintenance work is performed at least once a year.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.