MKF / MKFT (E5) 04/2022 page 122/176
21. Options
21.1 APT-COM™ 4 Multi Management Software (option)
The chamber is regularly equipped with an Ethernet interface (5) that can connect the BINDER APT-COM™
4 Multi Management Software. The MAC Address is indicated in the “Device info” controller menu (chap.
14.5.2.2). The actual temperature and humidity values are given at adjustable intervals. Programming can
be performed graphically via PC. Up to 100 chambers can be cross-linked. For further information on net-
working, please refer to the APT-COM™ 4 operating manual.
21.1.1 APT-COM™ 4 Basic Edition
APT-COM™ 4 Basic Edition is included with the chamber. APT-COM™ 4 is available for download on the
BINDER website. Upon registering the chamber, you will receive a license key with which you can activate
the functionality of the Basic Edition for your downloaded version.
Registration of the Multi Management Software APT-COM™ BASIC-Edition
Register now for getting your free BINDER Multi Management Software APT-COM™ 4 BASIC-Edition.
With the purchase of your BINDER chamber you will receive the BINDER Multi Management Software
APT-COM4™ 4 BASIC-Edition for free.
BINDER´s new Multi Management Software provides management, logging, programming and documen-
tation options and much more.
Important characteristics of APT-COM™ 4 BASIC-Edition:
Administration of up to five connected chambers
Log management (creating, deleting, archiving)
Documentation of recording values
Central overview of all chambers in both graphic and tabular form
Graphical presentation of recording values
Graphical/numerical program editor
Manual export of recording values (CSV/PDF file)
Multilingual user interface (German, English, French, Spanish, Italian)
Optional program execution via APT-COM™
Timer function
Import of data from APT-COM™ 3
Register your chamber today and request your personal software serial number.
Click here to register: https://www.binder-world.com/en/service-support/product-registration
21.2 RS485 interface (option)
With this option, the chamber is equipped with an additional 2-wire RS485 serial interface (5a) that can
connect the BINDER APT-COM™ 4 Multi Management Software. The actual temperature and humidity
values are given at adjustable intervals. For further information, please refer to the APT-COM™ 4 operating
manual.
21.3 Analog outputs for temperature and humidity (option)
With this option, the chamber is equipped with analogue outputs 4-20 mA for actual value and set-point
value of temperature and of humidity. These outputs allow transmitting data to external data registration
systems or devices.