MKF / MKFT (E5) 04/2022 page 80/176
Condition
Information
icon
Text information
Start after
condition occurred
Operation line “Compr. air dryer” on. Op-
tional compressed air dryer activated.
Turn operation line off when not using the
compressed air dryer!
“Compressed Air
Dryer”
immediately
Notifications are not shown in the event list.
11.1.2 Alarm messages
Condition Alarm message
Start after
condition occurred
Chamber door open “Door open after 5 minutes
Exceeded setpoint of the safety controller class 2 “Safety controller immediately
Exceeded maximum or minimum temperature (option
temperature safety device class 2)
“Temperature safety
device”
immediately
Temperature sensor defective
e.g. “- - - - or
“<-<-<” or “>->->”
immediately
Safety controller temperature sensor defective
“Safety controller
sensor”
immediately
MKF/MKFT from size 115 on:
Fault in refrigerating machine. Contact BINDER service.
Compressor
overcurrent
immediately
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
11.1.3 Messages concerning the humidity system
Alarm messages
Condition and measures Message
Start after condition
occurred
The humidity module is defective. Take the chamber out
of service and contact BINDER service.
The message must NOT be acknowledged!
“Humidity system” immediately
MKF 56:
The humidity module cannot fill up.
In case of freshwater supply via water pipe:
The water tap is closed, or the chamber is defective
(e.g. inlet valve of humidity module).
In case of freshwater supply via external freshwater can
(option):
Water can is empty. Humidification is turned off. In case
of refrigerating operation, the interior is strongly dehu-
midified. When the water supply is functional again, the
humidity system restarts, or the chamber is defective.
“Freshwater supply” immediately