MKF / MKFT (E5) 04/2022 page 81/176
Condition and measures Message
Start after condition
occurred
MKF/MKFT from size 115 on:
Freshwater can is too empty to allow normal function, or
the floating switch is defective.
The humidification system turns off.
Fill up the water can or open the water supply.
During the one-hour preheating phase: message without
significance.
“Freshwater can
empty”
after 60 sec.
MKF/MKFT from size 115 on:
Freshwater can is too full, or the floating switch is defec-
tive. Suck off the water (chap. 4.2.3). The chamber func-
tions as usual.
If the message persists, contact BINDER service.
“Freshwater can
overflow”
after 60 sec.
The humidity module cannot empty the condensate
tank.
Wastewater tube obstructed. Check the length and loca-
tion of the wastewater tube. If appropriate, contact
BINDER service.
or
Wastewater pump or floating switch of the wastewater
tank defective. Contact BINDER service.
“Wastewater” immediately
Notifications
Condition and measures Message
Start after condition
occurred
Maintenance of the humidity system is required. Contact
BINDER service.
“Humidity module
service”
after predefined
time (approx. 1
year)
Messages concerning the humidity system are shown in the event list.
When operating the chamber without water connection, turn off humidity control in the “set-
points” menu (chap. 6.3) in order to avoid humidity alarms.