21.8.2 Connecting the wastewater can
Hose connection
Plug the wastewater hose to the hose connection (23) of the wastewater can and secure it with a hose
clamp. You can use a part of the standard supplied water hose.
Hose connection (23)
of the wastewater can
Figure 26: Wastewater can
Plug the free hose edge to the wastewater connection “OUT” (14) at the rear of the chamber and secure it
with a hose clamp.
For emptying the wastewater can disconnect the hose first.
NOTICE
Danger of overflow of the wastewater can.
Damage to the surroundings of the chamber.
Regularly check the filling level of the wastewater can.
Always empty the wastewater can in a timely manner before it is full.
Bringing a source of humidity into the inner chamber may increase wastewater production.
Regularly check the filling level of the wastewater can.
21.8.3 Connecting with wastewater recycling
When the chamber interior is clean, you can reuse the wastewater from the chamber. Connect the
wastewater connection “OUT” (14) of the chamber with the hose connection (24) of the freshwater can. The
wastewater can is not used in this case.
NOTICE
Danger of soiling of the vapor humidification system.
Damage to the chamber.
Make sure to reuse wastewater ONLY with a clean chamber interior.
In case of soiling / contamination of the interior, conduct the wastewater to the
wastewater connection or use the wastewater can.