MKF / MKFT (E5) 04/2022 page 136/176
Condensation or icing at the
sides of the inner chamber.
Set-point for a long time below ambi-
ent temperature, icing in the preheat-
Defrost the chamber (chap. 18).
Condensation at the samples
or at the sides of the inner
chamber; information message
“Preheat phase” on the control-
ler display.
MKF 56: Main power switch (1)
turned on less than 1 hour before op-
Turn on the main power switch
(1) at least one hour before op-
MKF / MKFT from size 115 on: Main
power switch (1) and/or rear power
switch (12) turned on less than 1 hour
before operating the chamber.
Turn on the main power switch
(1) and/or the rear power switch
(12) at least one hour before
Control accuracy of +/- 2.5 %
Door opened very frequently. Open doors less frequently.
Humidity fluctuation, together
with temperature fluctuation
> 1 °C with a set-point approx.
3 °C above ambient tempera-
Place of installation too hot.
Select cooler place of installa-
tion or contact BINDER service.
No or low dehumidification.
Contact BINDER service.
Humidity control turned off.
Turn on humidity control (chap.
6.3, 7.3).
Alarm message “Humidity sys-
tem” on the controller display.
Humidity module is defective
Turn off the chamber and con-
tact BINDER service.
No dehumidification; infor-
mation message “Preheat
phase” on the controller dis-
play.
MKF 56: Main power switch (1)
turned on less than 1 hour before op-
Turn on the main power switch
(1) at least one hour before op-
MKF / MKFT from size 115 on: Main
power switch (1) and/or rear power
switch (12) turned on less than 1 hour
before operating the chamber.
Turn on the main power switch
(1) and/or the rear power switch
(12) at least one hour before
Icing at the sides of the inner
chamber.
Set-point was too long below ambient
temperature.
Defrost the chamber (chap. 18).
Condensation at the sides of
the inner chamber.
Combination of temperature/humidity
set-point values not in the optimum
range (see temperature humidity dia-
Select combination of tempera-
ture/humidity set-point values in
the optimum range (chap. 17).
Temperature set-point was too long
below ambient temperature, icing in
Defrost the chamber (chap. 18)
Combination of temperature/humidity
set-point values leads to falling below
Select suitable combination of
temperature/humidity set-point
Low humidity and temperature
accuracy
MKF 56: Fan speed has been re-
duced.
MKF 56: Set fan speed to
100%.
No chamber function
(dark display).
Display mode “Standby” active. Press on touchscreen.
Main power switch (1) not turned on
Turn on the main power switch
(1).
MKF/MKFT from size 115 on: Rear
power switch (12) not turned on.
Turn on the rear power switch
(12).