MKF / MKFT (E5) 04/2022 page 145/176
Chamber size
56
115
240
400
720
1020
Electrical data
Current type
1N~
3N~
3N~
3N~
3N~
3N~
Nominal power
kW
2.80
4.80
6.80
8.00
11.00
12.50
Power plug
IEC con-
nector
plug
(grounde
d)
CEE
plug
5-poles
16 A
CEE
plug
5-poles
16 A
CEE
plug
5-poles
16 A
CEE
plug
5-poles
32 A
CEE
plug
5-poles
32 A
Over-voltage category acc.
to
IEC
61010-1
II II II II II II
Pollution degree acc. to IEC 61010-1
2
2
2
2
2
2
Over-current release category B, internal
Amp
16
3 x 16
3 x 16
3 x 16
3 x 25
3 x 32
Different electrical data for chambers constructed for the USA and Canada
(model version MKF056-240V)
Nominal voltage (+/-10%) at 60 Hz
power frequency
V 240 -- -- -- -- --
Current type
2~
--
--
--
--
--
Electrical data of the voltage and frequency changer
System of protection acc. to EN 60529
IP
--
23
23
23
23
23
Nominal voltage (+/-10%) at 60 Hz
power frequency (input side)
V -- 480 480 480 480 480
Current type
--
3N~
3N~
3N~
3N~
3N~
Nominal power
kW
--
9
9
13
13
13
Over-voltage category acc.
to
IEC
61010-1
-- II II II II II
Pollution degree acc. to IEC 61010-1
--
2
2
2
2
2
Fuse
A
--
16
16
25
25
25
Environment-specific data
Noise level (mean value)
dB(A)
59
62
65
65
65
69
Noise level with optional compressed air
dryer (short-term) (mean value)
dB(A) 85 85 85 -- 85 --
Noise level with voltage and frequency
changer (mean value)
dB(A) -- 67 67 67 67 70
Energy consumption at +25 °C / 77 °F
and 60 % r.h.
Wh/h 800 1000 1500 1900 3000 2200
Filling weight of refrigerant R 452A
(GWP 2140)
kg / lbs.
1.50 /
33.1
2.00 /
4.41
2.20 /
4.85
3.10 /
6.83
5.00 /
11.02
5.00 /
11.02
Note: Chambers with voltage and frequency changer: Average heating up time reduced by 0.3 K/min each.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifica-
tion and may affect the humidity accuracy.

Table of Contents