MKF / MKFT (E5) 06/2020 page 144/176
The technical data refers to the defined usable volume.
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature and humidity.
25.4 MKF (E5) technical data
Width, gross (including 18 mm / 0.7 in for
1 access port (MKF 56, 115, 240), 36 mm
/ 1.4 in for 2 access ports (MKF 720) with
mm
inch
745
29.33
1000
39.37
1135
44.69
1615
63.58
Height, gross (incl. castors)
Depth, gross (incl. cable and door handle)
Depth, gross (incl. cable and door handle)
with optional compressed air dryer
Depth, gross (incl. cable and door handle)
with voltage and frequency changer
--
Wall clearance rear (minimum)
Wall clearance rear with optional com-
pressed air dryer or to set up the voltage
and frequency changer (minimum)
mm
inch
1000
39.37
1000
39.37
1000
39.37
1000
39.37
Wall clearance sides (minimum)
Width
Quantity of racks (regular)
Maximum permitted total load
Weight (empty) with optional compressed
air dryer
kg / lbs. 200 295 / 650 375 / 827 605 / 1334