Issue 03/2019 Art. no. 7001-0364
Operating Manual
Translation of the original operating manual
MK (E5)
Alternating climate chambers with program control
Model
Model version
Art. No.
MK 56 MK056-230V 9020-0374, 9120-0374
MK056-240V 9020-0388, 9120-0388
MK 115
MK115-400V
9020-0375, 9120-0375
MK115-400V-C 9020-0303 (with voltage and frequency changer)
MK 240
MK240-400V
9020-0376, 9120-0376
MK240-400V-C
9020-0355 (with voltage and frequency changer)
MK 720
MK720-400V
9020-0377, 9120-0377
MK720-400V-C
9020-0356 (with voltage and frequency changer)
MKT (E5)
Alternating climate chambers with deep temperature
and program control
Model
Model version
Art. No.
MKT 115
MKT115-400V
9020-0385, 9120-0385
MKT115-400V-C
9020-0363 (with voltage and frequency changer)
MKT 240
MKT240-400V
9020-0386, 9120-0386
MKT240-400V-C
9020-0364 (with voltage and frequency changer)
MKT 720
MKT720-400V
9020-0387, 9120-0387
MKT720-400V-C
9020-0365 (with voltage and frequency changer)
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • M
    Michelle FritzAug 17, 2025
    What to do if I need to dry and aerate my Binder Climate chamber after disinfection?
    • R
      Roger BrandtAug 17, 2025
      After using disinfectant spray in your Binder Climate chamber, allow the chamber to dry thoroughly, and ensure it is sufficiently aerated.
  • E
    Edward NewtonAug 20, 2025
    Why should I not empty disinfectant into drains when cleaning my Binder MK 56?
    • J
      James KellyAug 20, 2025
      Do NOT empty the disinfectant into drains to prevent environmental damage from disinfectant disposal.
  • A
    Andrew StoneAug 22, 2025
    How to troubleshoot a Binder climate chamber that has no function?
    • M
      Michelle HoodAug 22, 2025
      If your Binder climate chamber isn't working, first check the power supply connection. Then, verify that the power supply voltage is correct. Ensure the main power switch is turned on. For MK/MKT 115, 240, and 720 models, also check and turn on the rear power switch. Inspect the chamber fuse and replace it if needed; if it fails again, contact BINDER service.
  • V
    Victoria TaylorAug 26, 2025
    Why is there no or low refrigerating performance in my Binder climate chamber?
    • W
      William MillerAug 26, 2025
      If your Binder climate chamber has low refrigerating performance, ensure the ambient temperature is not higher than 25 °C / 77°F and if it is, select a cooler place for installation. If the problem persists, contact BINDER service to check the compressor.
  • S
    sandra00Aug 28, 2025
    What to do if Binder climate chamber is heating permanently and not maintaining the set-point?
    • C
      Carol WilliamsAug 28, 2025
      If your Binder climate chamber is constantly heating and not maintaining the set-point, calibrate and adjust the controller. If this doesn't resolve the issue, contact BINDER service.
  • E
    Elizabeth Johnston MDAug 30, 2025
    What to do if Binder climate chamber doesn’t heat up when turned on and safety controller responds?
    • M
      Megan MoranAug 31, 2025
      If your Binder climate chamber doesn’t heat up when turned on and the safety controller responds, acknowledge the alarm on the controller. Then, check the temperature setpoint setting. If it’s too low, select a suitable safety controller setpoint.
  • R
    Roberto HernandezSep 2, 2025
    Why my Binder climate chamber doesn’t heat up?
    • M
      Matthew SmithSep 2, 2025
      If your Binder climate chamber isn’t heating up, contact BINDER service to check the heating element.
  • M
    Michael CookSep 5, 2025
    Why does my Binder climate chamber display “Preheat phase” and has no refrigerating performance?
    • M
      Mr. Daniel ShepherdSep 5, 2025
      If your Binder climate chamber shows “Preheat phase” and has no refrigerating performance, ensure the main power switch (and rear power switch for MK / MKT 115, 240, 720 models) was turned on at least one hour before operating the chamber.
  • X
    xallisonSep 8, 2025
    What to do if acknowledging the alarm does not cancel the alarm state on Binder climate chamber?
    • R
      rgrimesSep 8, 2025
      If acknowledging the alarm on your Binder climate chamber does not cancel the alarm state, remove the cause of the alarm. If the alarm continues, contact BINDER service.
  • P
    Paul DominguezSep 12, 2025
    How to prevent condensation on samples in Binder climate chamber?
    • B
      browndawnSep 12, 2025
      If you notice condensation on the samples in your Binder climate chamber, use the anti-condensation protection feature.

Binder MK 56 Specifications