MK / MKT (E5) 03/2019 page 67/158
11. Notification and alarm functions
11.1 Notification and alarm messages overview
11.1.1 Notifications
Notifications are indicated by information icons displayed in the screen header in Normal display
An information icon serves as an indication of a certain condition.
If this condition persists, in some cases an alarm will be triggered after a fix or configurable interval. As
long as the condition persists, the information icon therefore continues to be displayed also in state of
alarm. If during alarm the conditions ends, e.g., if during a tolerance range alarm the actual value returns
to within the tolerance range, the information icon disappears, whereas the alarm will continue until manu-
al acknowledgement.
Press the flash icon next to the information icon to access the corresponding text information.
Normal display showing the text information.
The currently valid information texts are highlighted
in black (example: “Idle mode”)
Condition
Information
icon
Text information
Start after
condition occurred
1-hour preheating phase, no cooling function
“Preheating
phase”
immediately
Open chamber door
“Door open” immediately
The controller is in Idle mode (chap. 5.4).
“Idle mode immediately
Operation line “Anti-condensation” on. Anti-
condensation protection is activated.
“Anti-
condensation”
immediately
Operation line “Compr. air dryer” on. Optional
compressed air dryer activated.
Turn operation line off when not using the
compressed air dryer!
“Compressed Air
Dryer”
immediately
Notifications are not shown in the event list.