MK / MKT (E5) 03/2019 page 23/158
Do not install or operate the alternating climate chamber in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
4. Installation and connections
4.1 Connection of cooling water outlet for water cooling (option for MK 56)
An enclosure inside the chamber contains the connection kit for the cooling water inlet and outlet.
Fasten the cooling hose to the connection “OUT” (16) on the rear of the chamber (screw thread ¾’’).
You can use a part of the supplied tap water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against max. 50 °C / 122 °F.
Protect both ends of the drainage hose with two of the four supplied hose clamps. Before turning on
the chamber, check the connection for leaks.
4.2 Connection of cooling water inlet for water cooling (option for MK 56)
Connect the cooling water outlet before connecting the cooling water inlet.
Type of suitable water quality:
Water intake temperature: max. 10 °C / 50 °F.
pH value 4-7
connection pressure: 4 to 10 bar
BINDER GmbH is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.
Connection:
An enclosure inside the chamber contains the connection kit for the cooling water inlet and outlet.
Fasten the cooling water hose to the connection “IN” (17) on the rear of the chamber (screw thread
¾’’).
Install the water supply connection using either the enclosed water hose or another pressure-resistant
one. To accomplish this, remove the cover of the freshwater connection “IN” (17) on the rear of the
chamber.
The nominal diameter of the supplied water hose is ½ ", the length is 3m. The hose can be halved for
the inlet and the outlet.
The maximum hose length with nominal width ½ " is 5m. If a longer hose is necessary, a larger diame-
ter hose must be used.