1.6 Intended use
Alternating climate chambers MK / MKT are suitable for temperature treatment of solid or pulverized
charging material, as well as bulk material, using the supply of heat or cold. The chambers can be used
for drying processes, but they are particularly suitable for all tasks that occur in materials testing and aging
tests.
The chambers are suitable for harmless materials. A mixture of any component of the charging material
with air must NOT be explosive. The operating temperature must lie below the flash point or below the
sublimation point of the charging material. Any component of the charging material must NOT be able to
release toxic gases.
Other applications are not approved.
Do NOT use the chamber for drying purpose, especially if greater quantities of steam leading to conden-
sation will be set free.
The chambers are not classified as medical devices as defined by the Medical Device Directive
93/42/EEC.
Due to the special demands of the Medical Device Directive (MDD), these ovens are not qual-
ified for sterilization of medical devices as defined by the directive 93/42/EWG.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 21) are part of the intended use.
The chambers do not dispose of any measures of explosion protection.
DANGER
Explosion or implosion hazard.
Danger of poisoning.
Danger of death.
Do NOT introduce any substance combustible or explosive at working temperature into
the chamber, in particular no energy sources such as batteries or lithium-ion batteries
NO explosive dust or air-solvent mixture in the inner chamber.
Do NOT introduce any substance which could lead to release of toxic gases.
The charging material shall not contain any corrosive ingredients that may damage the ma-
chine components made of stainless steel, aluminum, and copper. Such ingredients include in
particular acids and halides. Any corrosive damage caused by such ingredients is excluded
from liability by BINDER GmbH.
In case of foreseeable use of the chamber there is no risk for the user through the integration of the
chamber into systems or by special environmental or operating conditions in the sense of EN 61010-
1:2010. For this, the intended use of the chamber and all its connections must be observed.

Table of Contents

Related product manuals