8) IBP (Invasive Blood Pressure) – Optional
- Measurement Range : -50 ~ 400mmHg
- Accuracy : ± 2mmHg
- Measurement Sensitivity : 1mmHg
- Alarm Range
IBP
[mmHg]
Display
High
Low
Systolic
40 ~ 295, OFF
OFF, 30 ~ 290
Diastolic
30 ~ 285, OFF
OFF, 25 ~ 280
Mean
35 ~ 290, OFF
OFF, 30 ~ 285
9) Temp
- Measurement Range : 0 ~ 50℃
- Measurement Sensitivity : 0.1℃
- Accuracy : ± 0.1℃ (25.0 ~ 50.0℃)
± 0.2℃ (0.0 ~ 24.9℃)
- Alarm range : High alarm limit 20.0℃ ~ 49.0℃
Low alarm limit 10.0℃ ~ 40.0
10) EtCO2 : End-Tidal CO2 – Optional
- Transducer Type: Mainstream/Sidestream CO
2
Sensor
- Range : 0~150mmHg(0~19.7%)
- Measurement Temperature : 0~45℃
- Accuracy:
- Alarm Range : High: 2~99mmHg , Low : 1~98mmHg
CO
2
Density
Accuracy
0~40mmHg
±2mmHg
41~70mmHg
±5% of reading
71~100mmHg
±8% of reading
101~150mmHg
±10% of reading