73
11) Multi Gas – Optional
- Transducer Type: Mainstream CO
2
Sensor
- Range(Accuracy) :
Co2 : 0~15% ±(0.2% +2% of reading)
N20 : 0~100% ±(2% +2% of reading)
HAL, ISO, ENF : 0 ~ 8% ±(0.15% +5% of reading)
SEV : 0~10% ±(0.15% +5% of reading)
DES : 0~22% ±(0.15% +5% of reading)
- Measurement Temperature : 0~45℃
- Alarm Range :
Co2 High: 1~15% , Low : 0~14%
N20 High: 1~100% , Low : 0~99%
HAL, ISO, ENF High: 1~8% , Low : 0~7%
SEV High: 1~10% , Low : 0~9%
DES High: 1~22% , Low : 1~21%
12) Cardiac Output - Optional
- Measurement Method : Thermodilution
- Measurement range : 0.1 ~ 20 L/min
- Repeatability : 2% or 0.2 L/min
- Injectate temperature range : 0 ~ 27
- Blood temperature range : 30 ~ 40
- Injectate volume: 3, 5, 10cc
- Catheter : Swan-Ganz Standard Thermodilution
Pulmonary Artery Catheter
(Edward’s Life Sciences)
13) Printer – Optional
- Print Method : Thermal
- Type : Internal
- Paper width : 58 mm
- Valid Width : 150 ~ 300mm (Change to 25mm unit)
- Wave Printer : 3CH (All Parameter Selection.)