Maintenance Page 34 of 42
Epoch 2 Instructions for Use 1771011 Revision A BioTek Instruments, Inc.
Clean the Touchscreen
Do not use:
Strong solvents, such as alcohol, acetone, ammonium chloride, methylene
chloride, and hydrocarbons, which can permanently damage the surface
of the touchscreen.
Fibrous materials, such as paper towels, which can scratch the
touchscreen. Over time, dirt particles and cleaning agents can get trapped
in the scratches.
Materials
Dish soap or other mild cleaner
Deionized or distilled water
Lint-free disposable towels
Protective gloves
Lab coat
Biohazard trash bags
Safety glasses
Surgical mask
Procedure
Do not spray or soak the touchscreen! This will cause damage.
1. Eject the plate carrier and then turn off and unplug the instrument.
2. Moisten a clean lint-free disposable towel with water, or with water and dish soap.
Wring out the towel so that liquid does not drip from it. Do not soak the towel.
3. Wipe the touchscreen gently with the moist towel.
4. If detergent was used, wipe the touchscreen with a towel moistened, not soaked,
with water.
5. Dry the screen gently using another towel.