Maintenance Page 33 of 42
Epoch 2 Instructions for Use 1771011 Revision A BioTek Instruments, Inc.
Maintenance
Schedule
Task
Frequency
Clean external surfaces
As needed
Inspect/clean touchscreen
As needed
Decontamination
Before shipment or storage
Clean Exposed Surfaces
Materials
Deionized or distilled water
Mild detergent
Clean, lint-free cotton cloths
Protective gloves
Lab coat
Biohazard trash bags
Safety glasses
Surgical mask
Procedure
1. Eject the plate carrier and then turn off and unplug the instrument.
2. Moisten a clean cotton cloth with water, or with water and mild detergent. Wring out
the cloth so that liquid does not drip from it. Do not soak the cloth.
3. Wipe the plate carrier, the inside of the plate carrier door, and all exposed surfaces of
the instrument.
4. If the reader is equipped with a cuvette holder, remove the holder and clean its
exposed surfaces.
5. If detergent was used, wipe all surfaces with a cloth moistened, not soaked, with
water.
6. Use a clean, dry, lint-free cloth to dry all wet surfaces.
If liquid is spilled inside the reader, contact BioTek Technical Assistance Center.