Specifications Page 39 of 42
Epoch 2 Instructions for Use 1771011 Revision A BioTek Instruments, Inc.
Specifications
General Specifications
Microplates
The Epoch 2 accommodates standard 6-, 12-, 24-, 48-, 96-, and 384-well microplates with 128 x 86
mm geometry and the Take3 Micro-Volume Plate. If using Gen5, the Take3 Trio Micro-Volume plate
and BioCell are also supported.
Maximum plate height: 20.32 mm
Hardware and Environmental
Weight:
Fully configured weight ~11.3 kg
Dimensions:
Touchscreen models: 31.8 cm x 32.4 cm x 39.4 cm
Non-touchscreen models: 21.6 cm x 32.4 cm x 39.4 cm
Operational Temperature
Range:
Touchscreen models: 18°C to 30°C
Non-touchscreen models: 18°C to 40°C
Humidity:
10% to 85% relative humidity (non-condensing)
Power Consumption:
Powered from an external 120W (minimum), 24VDC power supply
compatible with 100-240 volts AC @50-60Hz.
Temperature Control
Range:
Maximum incubation temperature 65°C. Minimum setpoint must be
ambient room temperature plus 4°C. For incubation setpoints > 60°C,
ambient room temperature must be at least 22°C.
Uniformity:
+/- 0.5°C at 37°C, tested with Innovative Instruments, Inc. temperature
test plate.