Specifications Page 40 of 42
Epoch 2 Instructions for Use 1771011 Revision A BioTek Instruments, Inc.
Absorbance Specifications
For the performance specifications described in this section, the gain on the optics test
should be < 8.
Unless indicated otherwise, the specifications in the table below apply to both
microplate and cuvette.
Wavelength Selection System:
2 photodiodes (measurement & reference) each for the
plate channel and the cuvette channel
< 5 nm (microplate & cuvette)
0.000–4.000 OD (microplate normal mode, cuvette)
0.000–3.000 OD (microplate sweep mode)
0.001 OD (touchscreen control only)
0.0001 OD (Gen5 control only)
0.2 nm (standard deviation)
Accuracy
Tested with certified neutral density glass
96-well plate
normal read speed
delay after plate movement = 100 ms
0.000 to 2.000 OD: ±1.0% ±0.010 OD
2.000 to 2.500 OD: ±3.0% ±0.010 OD
384-well plate
normal read speed
delay after plate movement = 100 ms
0.000 to 1.500 OD: ±2.0% ±0.010 OD
1.500 to 2.000 OD: ±5.0% ±0.010 OD
96- and 384-well plate
sweep read speed
0.000 to 1.000 OD: ±1.0% ±0.010 OD