Instrument Testing Page 38 of 42
Epoch 2 Instructions for Use 1771011 Revision A BioTek Instruments, Inc.
5. Click Start Test.
6. Place the Absorbance Test Plate on the microplate carrier, with well A1 in the
proper location.
7. Click OK to run the test.
8. When the test is complete, the results report appears. Scroll through the report;
every result should show “PASS.”