14: Run a System Test
Running a System Test will confirm that the reader is set up and running properly, or will
provide an error code if a problem has been detected.
1. Select
System > Diagnostics > Run System Test. If prompted to select a reader,
select the
Synergy HTX and click OK.
2. When the test is complete, a dialog will appear requesting additional information.
Enter the information (if required) and click
OK.
3. The results report will appear, with text that reads “SYSTEM TEST PASS.”
You may wish to print the report and store it with your Installation records.
The software stores system test information in its database; you can retrieve it
at any time.
If an error code is returned, turn to Appendix C, Error Codes and
look up the code. If the problem is something you can fix, do so now
and run another System Test. If the problem is something you cannot
fix, or if the test continues to fail, contact BioTek’s Technical
Assistance Center.
Dispenser 1 calibration
values
Dispenser 2 calibration
values
Synergy HTX Operator’s Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek Synergy HTX and is the answer not in the manual?

Biotek Synergy HTX Specifications

General IconGeneral
BrandBiotek
ModelSynergy HTX
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals