Chapter 8 Product Storage and Handling
Evia HF / HF-T Technical Manual
PAGE 91
8. Product Storage and Handling
8.1 Sterilization and Storage
The pulse generator is shipped in a cardboard box, equipped with a quality control seal and product
information label. The label contains the model specications, technical data, serial number, expiration
date, and sterilization and storage information of the pulse generator. The box contains a double
container with the pulse generator and product documentation.
The pulse generator and its accessories have been sealed in a container and gas sterilized with
ethylene oxide. To assure sterility, the container should be checked for integrity prior to opening. If a
breach of sterility is suspected, return the pulse generator to BIOTRONIK.
CAUTION
Storage (temperature) – Recommended storage temperature range
is -10° to 45°C (14°-113°F). Exposure to temperatures outside this
range may result in pulse generator malfunction.
Handling – Do not drop. If an unpackaged pulse generator is
dropped onto a hard surface, return it to BIOTRONIK.
CAUTION
FOR SINGLE USE ONLY – Do not resterilize the pulse generator or
accessories packaged with the pulse generator, they are intended for
one-time use.
Device Packaging – Do not use the device if the packaging is wet,
punctured, opened or damaged because the integrity of the sterile
packaging may be compromised. Return the device to BIOTRONIK.
Storage (magnets) – Store the device in a clean area, away from
magnets, kits containing magnets, and sources of electromagnetic
interference (EMI) to avoid damage to the device.
Use Before Date – Do not implant the device after the USE BEFORE
DATE because the device may have reduced longevity.
If a replacement pulse generator is needed, contact your local BIOTRONIK representative.