Chapter 5 Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 63
Mode Switch episode persisting beyond a programmable time limit
NOTE:
The attending physician must notify the BIOTRONIK Service Center about which of these events he/
she wishes to be informed.
5.18.5 Description of Transmitted Data
The following data are transmitted by the Home Monitoring system, when activated. In addition to the
medical data, the serial number of the implant is also transmitted.
The Monitoring Interval
The monitoring interval is considered the time period since the last periodic message was transmitted.
In a periodic report, the monitoring interval since the previous periodic report would be 24 hours.
The following data are transmitted for the Cardio Report by the Home Monitoring system, when
activated. In addition to the medical data, the serial number of the implant is also transmitted.
Device Status & Home Monitoring Settings
Containing device and message identifying values that pertain to the implant and Home Monitoring:
Implantation Date
Device Status
Remaining capacity for ERI calculation (done by the Service Center)
Last Follow-up
Message Creation Date/Time
Device Serial Number
Leads
Automatic Threshold Monitoring
Measured RV pacing threshold
LV enabled/disabled
RV enabled/disabled
RA enabled/disabled
Date/time of ATM measurement
Pacing Impedance (RA, RV, LV)
Sensing Amplitude (RA, RV, LV)
Pacing Counters (Brady)
AV-Sequences
Intrinsic Rhythm (AsVs)
Conducted Rhythm (AsVp)
Atrial Paced Rhythm (ApVs)
Complete Paced Rhythm (ApVp)
Atrial Arrhythmia
Atrial Tachy Episodes (36 out of 48 criteria)
Counter on AT/AF detections per day
Atrial Burden per day°
Ongoing Atrial Episode Time (programmable for 6, 12 or 18 hrs)