92
FFF
The time between the first falling edge of source X
and the first falling edge of source Y.
LRR
The time between the first rising edge of source 1
and the last rising edge of source 2.
LRF
The time between the first rising edge of source X
and the last falling edge of source Y.
LFR
The time between the first falling edge of source X
and the last rising edge of source Y.
LFF
The time between the first falling edge of source X
and the last falling edge of source Y.
If you want to measure voltage parameters, please follow the steps
below:
1. Press the MEASURE button to enter the Auto measurement
menu.
2. Press the first option button to enter the second measurement
menu.
3. Select measure type. If you press the “Voltage” option button,
“Voltage measurement” menu will display on the screen.
4. Press the Source option button to select CH1, CH2
according to input signal channel.
5. Press the Type option button to select the parameter type that
you want to measure. The corresponding icon and value will
display below the measure parameter.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 2190E and is the answer not in the manual?

BK Precision 2190E Specifications

General IconGeneral
BrandBK Precision
Model2190E
CategoryTest Equipment
LanguageEnglish

Related product manuals