Operation Modes 39
6.1.7 Short Mode
Short mode simulates a short-circuit at the input. In short mode, the DC load will draw the maximum
current from the source. The short function is available in any of the four operation modes (CC,CV,CW
or CR).
To select Short mode press the
+ buttons
The load will still report the previously selected mode. To stop short
mode press the + buttons once more. The load will return to its
previous operation. If CW mode was previously chosen the load will
not return to CW mode, the short current will continue to be drawn.
To stop the short, press the On/O key before pressing + .
When emulating a Short in CC, CW or CR mode, the maximum allowable short
current is equal to the 110% of current range