Conguration Menu 72
8.3 MEASURE
The MEASURE settings comprise parameters specically crafted to customize the process of capturing
voltage measurements. These settings encompass the voltage range, lter count, and voltage thresholds
for calculation of the voltage rise/fall time.
8.3.1 V-RANGE
The V-RANGE parameter enables/disables the auto voltage range function on a DC load. When enabled
the load will automatically adjusts the voltage range based on the applied load, ensuring optimal and
precise voltage measurements without manual intervention.
8.3.2 FILTER
Congures the measurement averaging lter. Elevating the averaging results in more precise readings,
albeit with a slower measurement update rate. Conversely, reducing the averaging leads to a faster
measurement update rate, albeit with less accuracy in the readings.
8.3.3 TIME-V1
Sets the start voltage threshold, determining the voltage level from which the timer for slew rate calculation
initiates.
8.3.4 TIME-V2
Sets the end voltage threshold, determining the voltage level from which the timer for slew rate calculation
terminates.
8.4 Sense
Enables/Disables the sense function. For more information regarding this function refer to Local/Remote
Sense.