99
Execution Errors
Error – 211: Trigger ignored
*TRG command was received, but it was ingnored because trigger subsystem
was not initiated (using INITiate command).
Error - 222: Data out of range
The numeric parameter value is out of range.
Error - 223: Too much data
A legal string program data element contains more data than the device could
handle due to memory or due to device specific requirements.
Error - 224: Illegal parameter data value
A legal type of parameter was received, but it is not the expected parameter for
the header.
Device Specific Errors
Error - 350: Queue overflow
Queue error is full, more than 20 errors occurred. No more errors will be saved
until error queue will be erased.
Error - 361: Parity error in program message
Parity bit not correct when data received over the RS-232 interface
Error – 362: Framing error in program message
Stop bit not detected when data received over the RS-232 interface (stop bit
detected as clear)
Error – 363: Input buffer overrun
Hardware input buffer of the serial port overflows with data because of the
improper reading of the buffer.
Error – 365: Time-out error