If one of the errors above is generated, the power supply must be turned off and
then turned on. If one of these errors is generated again, the power supply must
be delivered to B&K Precision for service.
Calibration Errors
Here are calibration error messages generated by the power supply:
Error 637: Secure code checksum failed
If this error is generated, the power supply must be unsecured using hardware
unsecure procedure. After that, the power supply must be secured. For more
information, see Calibration Overview section.
If the error persists, the power supply must be delivered to B&K Precision for
service.
Error 701: Calibration security disabled by jumper
The power supply was turned on and J6 jumper is set for calibrating mode
operation.
Error 703: Invalid secure code
The introduced secure code is not correct. (it does not match with the code
introduced when the power supply was secured)
Error 706: Secure code memory bad
If this error is generated, the power supply must be turned off and then turned
on. If the error persists, the power supply must be delivered to B&K Precision
for service.
Error 711: DAC offset register V start value out of range
Error 712: DAC offset register V out of range
Error 713: DAC V calibration failed
Error 714: DAC offset register V memory bad
Error 715: ADC offset register V out of range
Error 716: ADC offset register V memory bad
Error 721: DAC V step register start value out of range
Error 722: DAC V step register out of range
Error 723: DAC V step calibration failed