Supplemental Characteristics
Remote sensing capability
Voltage drop Up to 1V per each lead
Load regulation Add 5mV to spec for each 1-volt change in the + outputs
lead due to load current change.
Load voltage Substract voltage drop in load leads from specified output
voltage rating.
Temperature coefficient, ±(% of output+offset)
Maximum change in output/readback per
°
C after a 30-minute warm-up:
Voltage < 0.02%+2mV
Current < 0.1%+1mA
Output voltage overshoot
During turn-on or turn-off of AC power, output plus overshoot will not exceed
1V if the output control is set to less than 1V. If the output control is set to 1V
higher, there is no overshoot.
Programming language
SCPI (Standard Commands for Programmable Instruments)
State storage memory:
100 user configurable states
Recommended calibration interval:
1 year
Output terminal isolation (maximum, from chassis ground):
±60 Vdc when connecting shorting conductors without insulation between the
(+) output and the (+) sense terminals and between the (-) output and the (-)
sense terminals.
±240 Vdc when connecting insulated shorting conductors between the(+)
output and the (+) sense terminals and between the (-) output and the (-) sense
terminals.