67
Operations & Maintenance Manual: Product Specifications
Table 24. Accuracy Specifications
SPECIFICATION DESCRIPTION
Bladder volume accuracy ± (15% + 15 mL)
Bladder volume range 0999 mL (0–200 mL in small child scan mode)
The following example shows how the accuracy ranges shown in Table 24 could affect reported volume
measurements.
If the measurement scanned is 160 mL, with the accuracy range of ± (15% + 15 mL), the volume is calculated
as follows:
160 mL
×
15% = 24 mL
22.5 mL + 15 mL = 39 mL
150 mL ± 39 mL = 111189 mL
The accuracy specifications assume the instrument is being used according to the instructions provided by
Verathon
®
while scanning a tissue‑equivalent phantom.
Table 25. Operating & Storage Conditions
SPECIFICATION DESCRIPTION
Operating Conditions
Ambient temperature range 1040ºC (50–104ºF)
Atmospheric pressure range 7001060hPa
Relative humidity 30–75% non‑condensing
Ingress protection (IP) against
water
Console IPX0
Probe IPX1
Storage Conditions
Ambient temperature range 20–60ºC (‑4–140ºF)
Atmospheric pressure range 5001060hPa
Relative humidity 2095% non‑condensing