System Overview
1
Infusomat® Space 6.0 1 - 13EN
If the unit software version is not compatible with the Service
Program version, a window opens prompting the operator to
change the Service Program version. This window displays a
compatibility list of the Service Program- and unit software
versions.
If the software was updated or the processor PCB was re-
placed you are prompted to enter the device type. You can se-
lect either the Infusomat® Space P or the Infusomat®
S
pace S.
Note
The designation “Infusomat® Space S“ corresponds to the Infuso-
mat® Space.
The device type must be confirmed again.
If the Service Program and the unit software are
compatible,
all the Service Program functions are activated.
Fig.: 1 - 12
Fig.: 1 - 13
Fig.: 1 - 14

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1

Related product manuals