Disassembly / Assembly
3
3 - 14 Infusomat® Space 6.0
EN
Conversion, Retrofitting kits, and Spare Part Battery Contact
Strip
Designation Order No.
Retrofitting kit battery fastening . . . . . . . . . . . . . . . . . 3452 1365
sufficient for 10 devices
free delivery
Spare part battery contact strip . . . . . . . . . . . . . . . . . . 3452 1362
consisting of:
Battery contact strip with fixing hole
and connecting PCB
Conversion kit Infusomat® Space . . . . . . . . . . . . . . . . 3452 1361
solely for helicopter use
consisting of:
Battery contact strip with fixing hole
and connecting PCB
Housing upper part
Battery module with fixing pin
Battery compartment cover
Although our infusion pumps meet all the requirements for pro-
tection against vibrations
in accordance with EN 13718-1:2002,
our market observation has revealed sporadic wear of the battery
contacts on older Space infusion pumps. This applies only to de-
vices up to Serial Number 83672 in th
e original state whose hous-
ing has not been replaced yet.