Service Program
3
Perfusor® compact, 3.0 gb 3- 11
- Reloading free types
Select reloadable syringe types from the file list. Then
load the corresponding type. The data is displayed when
the type was loaded. Note the syringe type number and
the syringe type on the unit to ensure a clear assignment.
[Version 5.5004]: The article number and the syringe type
are only displayed after loading. Nothing is displayed
under “Version, free 20ml/50ml types”!
- Deleting free types
[Version 5.5004]: Reload the 20ml_0.spt or 50ml_0.spt
file with zero-value syringes. Now the corresponding free
syringe type is deleted.
[Version 2.1.0.1]: In window "Syringe," click
Delete
syringe
(Fig.: 3 - 11).
If necessary adapt the syringe selection (e.g. when the
selection was set to “Free type,” but all free types were
deleted).
Note
Adapt the syringe table after all modifications have been carried
out (see “Syringe Table and and Quick Reference Guide”
p. 4 - 5)
3.
Table ROM 20 ml
and
Table ROM 50 ml
Display of the syringe types saved in the Perfusor compact
ROM. The syringe types are numbered in sequence using the
syringe index [Version 2.1.0.1].
Modes Menu
1.
Modification data
Display and setting of:
max. basal rate, bolus rate, staff call, alarm tone, pressure
stage, last syringe type / index, Dianet address [Version
5.5004].
Alarm tone setting:
- [Version 5.5004]:
0 = continuous tone with 3 Hz intermittent
1 = continuous tone
- [Version 2.1.0.1]:
Continuous tone with 3 Hz intermittent
Continuous tone, static
Fig.: 3 - 12
[Version 5.5004]
[Version 2.1.0.1]
Fig.: 3 - 13
[Version 5.5004] [Version 2.1.0.1]
3.0