Service Program
3
3- 12 Perfusor® compact, 3.0 gb
Note
Please pay attention to the notes given with the staff call cable.
CAUTION
The pressure stage which was set last and the syringe type that
was selected last are overwritten with the pump settings when
the unit is switched off.
The values set are to be checked directly on the Perfusor®
compact when the maximum delivery rate, the bolus rate and
the syringe selection were changed and the Service Program
is quit.
CAUTION
A pressure calibration must be carried out if the bolus rate was
changed.
Constants Menu
1.
Service interval
Reads and resets the service interval timer. When the time set
has elapsed a service interval alarm is triggered when the unit
is switched on.
The timer can be set to 20440 hours (corresponds to an
average operation of 7 hours per day over 8 years). Counter
reading can be changed by overwriting [Version 2.1.0.1].
If this counter runs down to zero, a service alarm is triggered
every time the Perfusor® compact is switched on and a
service key flashes on the LC display. The audible alarm can
be acknowledged for the therapy time.
Note
Other menu items are of no importance to service.
Fig.: 3 - 14
[Version 2.1.0.1]
[Version 5.5004]
3.0