Service Program
3
3- 14 Perfusor® compact, 3.0 gb
Alarm Tone
1. Switch on unit.
2. Insert syringe and confirm (or select), e.g. OPS 50 ml.
3. Enter rate and actuate the Start button to start delivery.
4. Open syringe holder, an alarm is triggered.
5. Compare the alarm tone with the settings:
- [Version 5.5004]:
0 = continuous tone with 3 Hz intermittent
1 = continuous tone
- [Version 2.1.0.1]:
Continuous tone with 3 Hz intermittent
Continuous tone, static
Syringe / Syringe Selection
Note
Note down the current code for 20 ml and 50 ml syringes before
starting the test.
Test for table or internal setting:
1. Open holder, then press keys 7 C X X F and 7 C Y Y F
The syringe is accepted (if an alarm is triggered, then the
syringe selection is not correct!).
2. If a free type was input but is not accessible, then you should
check in addition:
Press keys 7 C {number of the 20 ml free type} F and keys 7
C {number of the 50 ml free type} F – alarm (if the syringe is
accepted, then the syringe selection setting is not correct!)
Test Code 20 ml Test Code 50 ml
Wildcard in the
subsequent text
XX YY
up to unit number
50920
21 51
from unit number
50921 on
24 61
Table 3 - 1
3.0