Service Program
3
Perfusor® compact, 2.1 gb 3- 15
2.1
Test for setting the free type or EEPROM:
3. Open holder, press keys 7 C {number of the 20 ml free type}
F and keys 7 C {number of the 50 ml free type} F – syringe is
accepted (if an alarm is triggered, then the syringe selection
setting is not correct!)
Press keys 7 C X X F and keys 7 C Y Y F – alarm (if the syringe
is accepted, then the syringe selection setting is not correct!)
Test for table and free type setting:
4. Open holder, press keys 7 C X X F and keys 7 C Y Y F – syringe
is accepted (if an alarm is triggered, then the syringe
selection setting is not correct!)
5. Open holder, press keys 7 C {number of the 20 ml free type}
F and keys 7 C {number of the 50 ml free type} F – syringe is
accepted (if an alarm is triggered, then the syringe selection
setting is not correct!)
Test for OPS setting:
6. Open holder, press keys 7 C 5 0 F and keys 7 C 2 0 F – syringe
is accepted (if an alarm is triggered, then the syringe
selection setting is not correct!)
7. Press keys 7 C X X F and keys 7 C Y Y F – alarm (if the syringe
is accepted, then the syringe selection setting is not correct!)
8. Reset syringe types after the test is finished!
Syringes 20 ml Free Type / Syringe Type Number
1. Open holder, press keys 7 C {number of the 20 ml free type}
F – syringe is accepted (if an alarm is triggered, then the
syringe selection setting is not correct!) Check whether the
free type was marked on the unit (please see syringe table).
2. Reset syringe types after the test is finished!